This is a video in the How To Calibrate Your Gas Monitor series by Ideal Calibrations.  All you need is trusted calibration gas (of course we recommend Ideal Gases brand), a cal cup, a regulator, and tubing.  

Industrial Scientific Ventis MX4 Gas Calibration Guide by Ideal Calibrations
1. Verify gas blend and expiration date.
2. Press and Hold Middle Button wait for screen lights up then release.
3. Press the left button 2 times.
4. Press the right button, and wait for zero to complete.
5. Press the right button again to begin calibration.

Calibration Gas for the Industrial Scientific Ventis MX6 is:
25ppm Hydrogen Sulfide
100ppm Carbon Monoxide
25% LEL Pentane (or 50% LEL Methane, which we recommend)
19.0% Oxygen
Nitrogen Balance

Buy the calibration gas here:
https://idealcalibrations.com/calibra...

Buy the Regulator here:
For Pumped (In video) - https://idealcalibrations.com/demand-...
For Diffusion (Non Pumped) - https://idealcalibrations.com/0-5-lpm...

Buy a replacement Oxygen Sensor here:
https://idealcalibrations.com/sensor-...

If you have any gas monitor or calibration gas questions, please feel free to call us at 1-734-956-0539 or to email me [email protected].  We'd be happy to answer any questions you have.

Please see our websites at:
http://www.idealcalibrations.com
12 سال پیش در تاریخ 1391/12/14 منتشر شده است.
100,609 بـار بازدید شده
... بیشتر